Portada » QA » PHmeters, conductivity meters, thermo-hygrometers and breathalyzers » Portable, pocket and laboratory digital pH meters and conductivity meters » Portable multiparameters
Multiparameter PC70-2
Price: €667 + VAT
Multiparameter PC70-3
Price: €677 + VAT
(includes pH electrode)
multiparametric-portable
pH, CONDUCTIVITY, TDS and Cº
Multiparameter with complete case that includes all the necessary accessories for the measurement of pH, Conductivity, Salinity, TDS and Temperature. Includes GLP function ( good laboratory practice ).
The backlit LCD screen simultaneously displays the current measurement of pH, conductivity and temperature, as well as the stability icon and the indication of the standards used in the calibration.
It incorporates the GLP function, backlit display with date and time, 500 memories with data with date and time and Cal Timer .
With USB cable, PClink software that manages the data output to the PC, manuals in Spanish and power supply.
Resistant to dust and water (IP57 protection), with a soft rubber cover that completely adapts to the shape of the equipment.
PC 70 GLP portable multiparameter
- complete suitcase
- pH electrode (PC70-3 model only)
- Conductivity cell with CAT (G-2301 TN)
- Cable for pH electrode and 1m fixed cable
- Standard solutions pH 4.01, 7.00
- Conductivity solutions 1413 μS, 12.88 mS
- USB cable + Pclink Software for data output to PC
- Spanish manual and mains adapter
PC 70 | ||
pH measurement scale | -2.00…16.00 pH | |
Resolution | 0.1/0.01pH | |
Calibration points | 1, 2 or 3 | |
Recognized buffers | USA: 1.68 – 4.01 – 7.00 – 10.01 pH NIST: 1.68 – 4.01 – 6.86 – 9.18 pH 2 user defined values | |
Asymmetry and Potential Indication | Yes with date and time |
Multiparameter Tab Tester | ||
Establish measured stability criteria | Yes | |
LPG function | Yes | |
Calibration time | Yes | |
Indication of used buffers in calibration | Yes | |
mV Measurement scale | ±1,999mV |
Multiparameter Tab Tester | ||
Resolution | 0.1mV (±200mV) / 1mV | |
COND Measurement scale | 0.0…199.9mS | |
Resolution | 0.01/0.1/1µS 0.01/0.1mS | |
Calibration points | 1…4 | |
Recognized buffers | 84µS – 1413µS – 12.88mS – 111.9mS 1 user-defined values | |
cell constant | 0.1 – 1 – 10cm- 1 | |
temperature coefficient | 0.00…10.00%/°C | |
reference temperature | 15…30°C | |
LPG function | Yes | |
Timer Calibration | Yes | |
Indication of used buffers in calibration | Yes | |
TDS measurement scale | 0…100 g/l | |
Resolution | 1% fs | |
TDS factor | 0.4…1.0 | |
temperature measurement scale | -10…110 °C | |
Resolution/Accuracy | 0.1°C / ±0.5°C | |
pH temperature compensation | Automatic/manual 0…100 °C | |
COND temperature compensation | Automatic and manual 0…100ºC | |
System: LPG | Yes | |
Memory | Man / Auto 500 data with date and time | |
auto off | After 20 min. | |
Screen | LCD (Liquid Crystal Display) backlit | |
Tickets | BNC, phono jack (CAT) and USB | |
Feeding | 3 x 1.5V AAA batteries AC/DC adapter with USB cable | |
battery life | > 300 hours | |
IP degree of protection | WaterproofIP57 | |
Instrument dimensions and weight | 86 × 196 × 33mm / 300g | |
Suitcase dimensions and weight | 385 x 300 x 115mm / 1725g |
>Multiparametric PC7-2
Price: €567 + VAT
Multiparameter PC7-3
Price: €678 + VAT
(includes pH electrode)
Multiparameter PC7
pH, CONDUCTIVITY, TDS and Cº
The PC7 They satisfy the demands of a robust and simple equipment, designed for hard work, although they do not reach all the features of the PC70.
PC 7 portable multiparameter
- complete suitcase
- pH electrode (PC70-3 model only)
- Conductivity cell with CAT (G-2301 TN)
- Cable for pH electrode and 1m fixed cable
- Standard solutions pH 4.01, 7.00
- Conductivity solutions 1413 μS, 12.88 mS
- Accessories, Spanish manual and mains adapter
pcs 7 | |||
pH measurement scale | 0.00…14.00pH | ||
Resolution | 0.1/0.01pH | ||
Calibration points | 1, 2 or 3 | ||
Recognized buffers | USA: 1.68 – 4.01 – 7.00 – 10.01 pH NIST: 1.68 – 4.01 – 6.86 – 9.18 pH 2 user defined values | ||
Asymmetry and Potential Indication | Yes | ||
Establish measured stability criteria | Yes | ||
LPG function | – | ||
Calibration time | – | ||
Indication of used buffers in calibration | Yes | ||
mV Measurement scale | ±1,000mV | ||
Resolution | 1mV | ||
COND Measurement scale | 0.0…199.9mS | ||
Resolution | 0.1/1µS 0.01/0.1mS | ||
Calibration points | 1…4 | ||
Recognized buffers | 84µS – 1413µS – 12.88mS – 111.9mS 1 user-defined values | ||
cell constant | 0.1 – 1 – 10cm- 1 | ||
temperature coefficient | 0.00…10.00%/°C | ||
reference temperature | 15…30°C | ||
LPG function | – | ||
Timer Calibration | – | ||
Indication of used buffers in calibration | Yes | ||
TDS measurement scale | – | ||
Resolution | – | ||
TDS factor | – | ||
temperature measurement scale | 0….100.0 °C | ||
Resolution/Accuracy | 0.1°C / ±0.5°C | ||
pH temperature compensation | Automatic/manual 0…100 °C | ||
COND temperature compensation | Automatic and manual 0…80ºC | ||
System: LPG | – | ||
Memory | – | ||
auto off | After 20 min. | ||
Screen | LCD (Liquid Crystal Display) | ||
Tickets | BNC and phono jack (CAT) | ||
Feeding | 3 x 1.5V AAA batteries | ||
battery life | > 300 hours | ||
IP degree of protection | WaterproofIP57 | ||
Instrument dimensions and weight | 86 × 196 × 33mm / 295g | ||
Suitcase dimensions and weight | 385 x 300 x 115mm / 1720g |
pcs 7 pH, CONDUCTIVITY and Cº | PC 70 pH, CONDUCTIVITY, TDS and Cº | |||
pH measurement scale | 0.00…14.00pH | -2.00…16.00 pH | ||
Resolution | 0.1/0.01pH | 0.1/0.01pH | ||
Calibration points | 1, 2 or 3 | |||
Recognized buffers | USA: 1.68 – 4.01 – 7.00 – 10.01 pH NIST: 1.68 – 4.01 – 6.86 – 9.18 pH 2 user defined values | |||
Asymmetry and Potential Indication | Yes | Yes with date and time | ||
Establish measured stability criteria | Yes | Yes | ||
LPG function | – | Yes | ||
Calibration time | – | Yes | ||
Indication of used buffers in calibration | Yes | Yes | ||
mV Measurement scale | ±1,000mV | ±1,999mV | ||
Resolution | 1mV | 0.1mV (±200mV) / 1mV | ||
COND Measurement scale | 0.0…199.9mS | 0.0…199.9mS | ||
Resolution | 0.1/1µS 0.01/0.1mS | 0.01/0.1/1µS 0.01/0.1mS | ||
Calibration points | 1…4 | |||
Recognized buffers | 84µS – 1413µS – 12.88mS – 111.9mS 1 user-defined values | |||
cell constant | 0.1 – 1 – 10cm- 1 | |||
temperature coefficient | 0.00…10.00%/°C | |||
reference temperature | 15…30°C | |||
LPG function | – | Yes | ||
Timer Calibration | – | Yes | ||
Indication of used buffers in calibration | Yes | Yes | ||
TDS measurement scale | – | 0…100 g/l | ||
Resolution | – | 1% fs | ||
TDS factor | – | 0.4…1.0 | ||
temperature measurement scale | 0….100.0 °C | -10…110 °C | ||
Resolution/Accuracy | 0.1°C / ±0.5°C | |||
pH temperature compensation | Automatic/manual 0…100 °C | |||
COND temperature compensation | Automatic and manual 0…80ºC | Automatic and manual 0…100ºC | ||
System: LPG | – | Yes | ||
Memory | – | Man / Auto 500 data with date and time | ||
auto off | After 20 min. | |||
Screen | LCD (Liquid Crystal Display) | LCD (Liquid Crystal Display) backlit | ||
Tickets | BNC and phono jack (CAT) | BNC, phono jack (CAT) and USB | ||
Feeding | 3 x 1.5V AAA batteries | 3 x 1.5V AAA batteries AC/DC adapter with USB cable | ||
battery life | > 300 hours | |||
IP degree of protection | WaterproofIP57 | |||
Instrument dimensions and weight | 86 × 196 × 33mm / 295g | 86 × 196 × 33mm / 300g | ||
Suitcase dimensions and weight | 385 x 300 x 115mm / 1720g | 385 x 300 x 115mm / 1725g |